Abstract
An advanced-type small, light, multi-functional electronic personal dosemeter has been developed using silicon semiconductor radiation detectors for dose management of workers at nuclear power plants and accelerator facilities. This dosemeter is 62 × 82 × 27 mm3 in size and ∼130 g in weight, which is capable of measuring personal gamma ray and neutron dose equivalents, Hp(10), simultaneously. The neutron dose equivalent can be obtained using two types of silicon semiconductors: a slow-neutron sensor (1 MeV). The slow neutron sensor is a 10 × 10 mm2 p-type silicon on which a natural boron layer is deposited around an aluminium electrode. The fast neutron sensor is also a 10 × 10 mm2 p-type silicon crystal on which an amorphous silicon hydride is deposited. The neutron energy response corresponding to the fluence-to-dose-equivalent conversion coefficient given by ICRP Publication 74 has been evaluated using a monoenergetic neutron source from 250 keV to 15 MeV at the Fast Neutron Laboratory of Tohoku University. As the result, the Hp(10) response to neutrons in the energy range of 250 keV and 4.4 MeV within ±50% difference has been obtained.

Publication details
Download http://rpd.oxfordjournals.org/cgi/content/short/126/1-4/284
http://dx.doi.org/10.1093/rpd/ncm059
Publisher Oxford University Press
Repository HighWire Press OAI Repository (United States)
Keywords POSTER Presentations
Type TEXT
Language English